OBJECTIVE: To study the influence that the scan percentage tool used in partial k-space acquisition has on the quality of images obtained with magnetic resonance imaging equipment.
MATERIALS AND METHODS: A Philips 1.5 T magnetic resonance imaging scanner was used in order to obtain phantom images for quality control tests and images of the knee of an adult male.
RESULTS: There were no significant variations in the uniformity and signal-to-noise ratios with the phantom images. However, analysis of the high-contrast spatial resolution revealed significant degradation when scan percentages of 70% and 85% were used in the acquisition of T1- and T2-weighted images, respectively. There was significant degradation when a scan percentage of 25% was used in T1- and T2-weighted
in vivo images (
p < 0.01 for both).
CONCLUSION: The use of tools that limit the k-space is not recommended without knowledge of their effect on image quality.
Keywords: Magnetic resonance imaging; K-space; Quality control; In vivo magnetic resonance imaging.