ORIGINAL ARTICLE
|
|
Comparative study of image quality and entrance and exit air kerma measurements on chest phantom utilizing analog and CR digital imaging systems |
|
Autho(rs): Renata Matos da Luz, Gabriela Hoff |
|
Keywords: Computed radiography, Image quality, Air kerma |
|
Abstract:
IGraduation, Medical Physics, Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS) - Faculdade de Física, Porto Alegre, RS, Brazil
INTRODUCTION In 1980, Fuji Film® from Japan developed a two-dimensional detector called imaging plate for medical applications, with the objective of replacing the traditional radiographic film. Since then, the need to reduce radiation dose while enhancing contrast with low noise levels has encouraged the research for new storage systems for latent diagnostic images. The substitution of radiography with image intensifiers by computed systems brought many operational advantages, such as the electronic images storage and image post processing. In the field of computed radiography, the concept of diagnostic image quality acquires a new meaning and interpretation due to the unique and dynamic attributes of the images generated in this system, with the possibility of manipulating images with the preprocessing of the digital signal. However, this capability may actually hide technical mishaps and cause a dose increase in the patient for a certain technique. Thus the need emerged to acquire new knowledge, concepts and to change existing procedures; that is to say that an entirely new set of problems is waiting for solutions. The patient absorbed dose is a variable to be managed in each system, as the manipulation of generated images may mask excessive doses. Additionally, the different technologies of the currently available digitized systems have highly differentiated signal conversion efficiencies and resolutions and may bring about a dose increase (intrinsic to each technology) particularly as compared with a film-based system. Because of such these differences intrinsic to the digital imaging technology, an appropriate calibration of these systems is critical to obtain the lowest dose required to achieve diagnostic image quality and thus benefiting from the advantages of this technology. Some authors, among them Souza et al.(1), have published studies on this subject. The present study was idealized and developed during the period of migration from the analog to the CR digital imaging system, with the objective of analyzing and comparing images quality and measurements of entrance and exit air kerma in a non-anthropomorphic chest phantom during procedures performed with both imaging systems. It also proposes a simple comparative method between different images that can be adapted to calibrate the digital systems in the migration process, as well as comparing both systems regarding image quality and external dosimetry.
MATERIALS AND METHODS Two different X-ray apparatuses, both at the same hospital in the city of Porto Alegre, RS, Brazil, were utilized during the transition from the analog to the computed radiology (CR) imaging system. One of the units is a Siemens Iconos model, which comprises a generator with a high-frequency wave rectifier, and the other is a non identified model manufactured by Emic, with a generator with a medium frequency wave rectifier (12 pulses). The phases in the data acquisition process were the following: assessment of equipment and instruments; checking of quality tests of the equipment; image acquisition following parameters recommended in the adopted literature(2); digitization of all obtained images to carry out the quality parameter calculations; study of the catalogued and emitted mean spectra for the characteristics of the X-ray equipment and their implications on the image quality and absorbed dose. The quality control tests performed demonstrated that the equipment was in compliance with Brazilian standards and appropriately operational. The analog image processing system comprised a model MX-2 Macrotec processor, whose standard development temperature was 33.4°C. The images were acquired one day after flushing and substitution of development chemicals utilized in such processor. The cassette utilized was manufactured by Konex, with 35 × 43 cm2 in size. Films and development chemical solutions were manufactured by Kodak. The digitized system, by its turn, was comprised by a CR 950 Kodak multicassette system, a Kodak DryView 8900 laser imaging system with Kodak films, and a Kodak cassette with imaging plate measuring 35 × 43 cm2. Images acquisition was performed with a polymethylmetacrilate (PMMA) non-anthropomorphic chest phantom measuring 30 × 30 × 20 cm3. The PMMA phantom has the thickness to present an optical density similar to that of a standard adult patient. An aluminum bar (Al) measuring 30.0 × 2.0 × 1.0 cm3 was added to the phantom in an appropriate slot provided for such purpose in order to simulate absorption and scattering by the spine. The images were acquired with both X-ray systems with the same exposure factors, equal focus-film distance and the same positioning adopted for chest radiography, as established in the usual literature on radiographic techniques(2). The exposure factors were varied until the generated images acquired a similar visual contrast, as follows: 102 kVp and 2 mAs, 102 kVp and 4 mAs and 117 kVp and 4 mAs, all of them with 200 mA anodic current (gross focus) at a distance of 1.5 m of the film focal point. With such exposure factors, three images were generated with the analog system and three others in the CR digital system, for each one of the apparatuses. With a view to the clinical applications of the present study, based on the principles of radiological protection and the application of the ALARA principle, it is not possible to separate the patient absorbed dose from the radiographic technique during the radiological procedure. In other words, an application to evaluate and/or improve the diagnostic quality of the image will only be valid when the dose to which the patient is exposed, is the lowest possible, while maintaining the quality criteria for diagnostic imaging. For this reason, exposure measurements were performed during procedures simulations with a duly calibrated Radcal 9010 ionization chamber. The chamber was aligned with the focal point on the acquisition geometry. Two measurements were performed, one at the primary beam entrance surface, and the other at the primary beam exit surface of the phantom. This way, the phantom entrance and exit air kerma were determined. The films containing the acquired images were digitized on a Microtek, Scan-Maker 1000 XL scanner specific for films. The images were digitized in 16 bits and 300 dpi with minimum parameters for contrast, brightness, enhancement and shadows, so that no image characteristics were altered. After digitization, all the images were converted into the DICOM format (Digital Imaging Communications in Medicine). The images acquired with the CR digital system were saved and processed without addition of any filters, by the Kodak system. Quality parameters represent a way to evaluate digital images quality(3). The main factors affecting digital image quality include: resolution, signal intensity, contrast, noise and signal-noise ratio. The values of these descriptors were obtained for each image by means of the Java public domain software ImageJ(4) for the development of applications for image processing and analysis. With this software, after selection of a region of interest (ROI), the area, mean, and standard deviation (SD) of the signal were obtained, as they play an essential role in the image quality parameters calculation. An approximately constant circular ROI was defined for all the images, over the simulated pulmonary field and over the region of the spine simulation. The contrast in the radiographic image is defined as the signal difference between two adjacent regions on the film. It is directly dependent upon the subject contrast(a) (irradiated material), reflecting differences between atomic numbers, electrons density, and thickness of the material of which the image is generated, as well as the characteristics of the energy spectrum of the emitted radiation beam(5-7). As an example, the measurement of radiation flow (F) between two adjacent regions of a patient, F1 and F2, such flows being different because of distinctive tissues constitution. Thus, the image contrast(b) (IC) may be measured as the difference in signal intensity (S), where S represents the mean ROI value, between two areas on the film or on the digital image, divided by the signal from the ROI area of interest , as shown in equation 1. Noise is the main factor limiting the digital image contrast; it is a signal that interferes in homogeneity, reducing the contrast. It is frequently defined as uncertainty in a signal due to random fluctuations of such signal(8,10). Possible causes for such fluctuations are originated in the statistical nature of the radiation beam emerging from the tube, whose number of photons emitted by time unit varies either as a Gaussian or Poisson distribution(10). Other random fluctuation sources are introduced by the process of radiation attenuation by mater, which also follows a Poisson distribution, besides the fact that the electronic system introduces noise because of the electronic binary processing(10). The SD is the most appropriate variable to characterize the noise in an image system, as it quantifies an image pixel values variation around a mean value(10), being determined for a Poisson distribution, according to equation 2. In the present study, the ROI SD was utilized as being representative of the image noise, besides the above estimate presented on equation 2. Based on the distribution of the signal intensity frequency chart, the SD was calculated as being the uncertainty of the values of signal intensity around a mean value. Equation 2 represents the mean number of pixels at the ROI. In order for the noise to have a physical meaning, it is associated with a signal at the ROI, so that it is utilized in the definition of the signal noise ratio (SNR). SNR describes a relation between the mean signal value and the variation around this mean value. By means of this descriptor, it is possible to define the efficiency of the image acquisition system. Equation 3 demonstrates the SNR. The images were also analyzed by means of their histograms, which are graphical representations of the frequency distribution as a function of the signal(3). They are statistical functions of the image, which at each level of gray scale, calculates the number of pixels in that particular gray scale level, so that the image contrast is determined by the window width and by the brightness level on the histogram. The increase in the window width decreases the contrast, but increases the possibility of representing a higher number of gray tones that can be seen by the "naked eye", while the increase in level increases the brightness(10).
RESULTS The results comprise data that were evaluated based on the image characteristics (Table 1) and on collected air kerma data. The CR digital system images that presented best visual contrast and have visually similar characteristics for the two types of equipment were those acquired with exposure factor of 117 kVp with 4 mAs. For that reason, images obtained under such technical conditions were analyzed in particular. Figures(c) 1 and 2 compare the images obtained in the two apparatuses, for both analyzed systems.
Another point to be considered, besides patient absorbed dose, is the fact that in order to generate an image in the studied CR digital imaging system, a greater photons flow was necessary as compared with the analog system for a given exposure index. An analysis of the photon spectra may assist in the understanding of the differences between the two apparatuses and the different responses for each image transducer. For this purpose, the charts presented on Figures 3 and 4 show, respectively, the X-ray equipment emission spectrum, considering a tube added of a tungsten track operating at 90 kVp, angle of electron flight of 22° and a 3 mm-thick Al filter, for a ripple of 5% and 15%, and the different filtrations for spectra with different ripples(11).
DISCUSSION Based on the data presented on Table 1, considering the images generated on the Siemens equipment with the analog system, the signal was reduced from 15.6 to 3.0 after the increase in the tube voltage from 102 kVp to 117 kVp, with a constant tube current at 4 mAs. In this system, with the increase in voltage and in transported load, a signal reduction of 527% of the initial value was observed. For the CR digital imaging system, the signal value was 114% lower for the voltage of 117 kVp as compared with the lowest voltage of 102 kVp, with the highest signal value being 35.9. On the other hand, with respect to the images generated with the Emic equipment, the analog processing resulted in a signal increase of 247% with the duplication of the load transported in the tube, since for the factors 102 kVp and 2 mAs the signal value was 6.4, while for 102 kVp and 4 mAs that value was 15.8, an increase of 186%, with the tube voltage increase. The signal values for the exposure factors of 102 kVp and 2 mAs and 117 kVp and 4 mAs were, respectively, 6.4 and 29.4. For the Emic equipment and the CR digital imaging system, the signal increased 107%, with the duplication of mAs, where the signal value for 104 kVp and 2 mAs was 35.5, and for 102 kVp and 4 mAs, was 37.9. A reduction of 112% of the initial signal value was observed when the tube voltage was increased from 102 kVp to 117 kVp. Thus, in the case of analog image processing, SNR increased with transported load and was reduced with the increase in voltage. On the other hand, in the CR signal transducer system, SNR increased both with voltage increase and current increase in the tube. The images on Figures 1 and 2 will help in the discussion of results presented on Table 1. The analysis of data regarding signal intensity obtained from images generated in the Siemens equipment did not present significant changes when mAs was increased from 2 mAs to 4 mAs, with constant 102 kVp, independently of analog or digital processing. However, when voltage was increased from 102 kVp to 117 kVp, in both cases there was a decrease in signal intensity. The images generated by the analog system are over exposed and present low contrast (darkened image), an evidence of the excessive radiation captured by the transducer system. The ideal image for this system is obtained at 102 kVp and 2 mAs for both apparatuses. On the other hand, the images generated by the CR digital imaging system presented low contrast, and according to the medical staff at the hospital, even with the digital processing resources, the diagnostic quality could not be attained. This characteristic was particularly more evident in the Emic equipment with 12 pulse wave rectification. Based on data presented on Figures 1 and 2 and data on Table 1, it is possible to observe that an image contrast reduction occurs in the analog image generated by the Siemens equipment (high frequency rectification). This is caused by the fact that spectrum generated by the Siemens equipment has, on average, a higher energy level than that of the Emic equipment, as for the same exposure technique, the visual result was significantly different for this imaging system. Furthermore, the signal is reduced with the increase in voltage in the high frequency equipment, while in the equipment with 12 pulse wave rectification, the signal increases. This directly affects the image contrast, which is lower in the high frequency equipment. Data on Table 1 show that the difference between the digital images resides mainly in the increase in image noise, represented on Figure 2B. This is clearly observed on the dotting of the image and in the widening of gray scale of the images histogram, which demonstrate the noise representation on such images. Usually, the histogram with a peak in the signal intensity distribution at the center of the gray scale is observed in standardized and stable processes, in which the quality characteristic is continuous. With the obtained air kerma results, the authors could observe that phantom exit flow for the Emic equipment, with medium frequency wave rectification (12-pulse rectification and ripple of approximately 15%) is lower than that obtained with the Siemens equipment, that has high-frequency wave rectification (ripple < 4%). Additionally, the spectrum generated by the Emic equipment presents lower mean energy than that generated by the Siemens equipment, which is related to wave rectification differences between both apparatuses. This characteristic has a direct influence on the quality of the X-ray spectrum, resulting, for the Emic equipment, in a spectrum with lower penetration in the phantom and, consequently, in images with lower definition and greater noise. This is caused by the decrease in the phantom exit flow. In order to obtain an image similar to those from the Siemens equipment, a different technique would be required, with greater exposure factors, and consequently, increased patient absorbed dose. The value calculated for the ratio between phantom entry and exit flow for the Emic equipment is greater than the ratio with the Siemens equipment. This happens because the mean energy of the spectrum emitted by the Emic equipment (Figure 3) is lower than that emitted by the Siemens equipment and, as the beam penetrates the phantom, a greater energy deposition occurs on its different layers, obviously depending on the thickness and composition of the phantom, which results in different exit flows (Figure 3). Comparing the spectra of the studied equipment, the Siemens one always presents a greater photons flow in the same exposure conditions and phantom thickness, if one takes into account only the possible spectrum generated by each equipment and its alterations caused by each system's ripple. In the Emic equipment, with spectra that presented lower penetration power than that of the Siemens equipment, there will be added wear of the X-ray tube, as it will operate at higher temperatures, which may damage the anode. In other words, many times the compensation will be performed by the automatic exposure control, increasing the mAs. Actually, this increase in mAs (usually in exposure time) generates an increase in the patient absorbed dose and reduces the useful tube life because of the anode heating and wear. Considering the studied irradiation geometry and the X-ray beam transport from the focal point, as the phantom is penetrated, the beam may only pass through 20 cm of PMMA, or by 18 cm of PMMA (representing the entry flow on the aluminum bar in the phantom), or pass through the 18 cm of PMMA and the 2 cm Al bar (representing the phantom exit flow). Figure 4(A,B) exemplifies the spectrum alterations when passing through the phantom. For each studied spectrum, 90 kVp spectra were utilized as, in the catalog, these were the closest spectra to the ones studied, considering the characteristic equipment's ripple changes. As Figures 4A and 4B are compared, one can realize that the final photons flow after crossing the material is greater for the 5% ripple, indicating that for a 15% ripple, the energy and, consequently, the spectrum penetration power is smaller. This directly influences the generation of quantum noise on the image, this spectrum being not advisable for use with transducers of low conversion efficiency. The reduction in the area under the curves is quite visible on the chart representing the spectrum with 15% ripple in comparison with the spectrum with 5% ripple in all studied cases. This analysis becomes important as the two different spectra undergo different alterations before reaching the transducer, with different flows that will be differently recorded, influencing the image quality and the dose in the procedure. Actually the present study was developed after several tube changes due to burn-outs in the Emic equipment (with 12-pulse rectification). The present study corroborates the indication that, when migrating from analog to digital system, it is important to consider the conversion efficiency of the X-radiation of the latent image from the transducer and the type of equipment available (specially the wave rectification in the input generator). In case of systems with 12-pulse wave rectification, or ripple of > 15%, the transducer conversion efficiency should be evaluated to assure the image quality and the equipment integrity.
CONCLUSIONS The measured air kerma values demonstrated that the equipment with wave rectification presenting a ripple > 15% generates a significantly smaller photons flow at the phantom exit than that of the equipment with wave rectification with a ripple < 5%. On the other hand, the image quality parameters were more appropriate on the images generated by the CR digital imaging system, when wave rectification systems with ripple of < 5% were utilized. When the equipment with wave rectification presenting ripple of > 15% was utilized, the determined exposure level had to be higher than that recommended by the literature(8), in order to reduce noise on images, which would otherwise cause the image to present a low diagnostic quality. Therefore, it is important to determine the most appropriate radiographic technique for the CR digital imaging system, in order to avoid an increase in dose to the patient and a decrease in the diagnostic image quality. In the present study, the image with the best characteristics for diagnosis was generated in the Siemens equipment with 117 kVp and 4 mAs, at 1.5 m from the focal point. It is important to note that the appropriate radiographic system should be determined for each imaging system and operating team. The contribution of the present study is the determination of a relatively simple and reproducible method for adapting the radiographic technique by means of the comparison of dose and image descriptors, and/or a method that can be utilized for comparing digitized image apparatuses from different institutions with a low cost, given the instruments utilized. It is important to observe that the proposed process is only valid if the image descriptors criteria are in agreement with the minimum image quality required for diagnosis. At the present stage, the authors suggest that whenever an institution plans on the acquisition of a CR digital imaging system, an evaluation of its X-ray systems characteristics be carried out, considering the photons flow and signal conversion efficiency required for the image generation in the CR digital imaging system. Otherwise, the acquisition of new X-ray equipment may be necessary after the deployment of the CR digital imaging system, which will bring an additional and unforeseen cost. Additionally, the utilization of reference dose levels defined by the Order (Portaria) No. 453(12) of the Brazilian Ministry of Health is suggested, for the examinations to be performed in the recently installed systems, in order to assure that the dose to the patient is the lowest possible(1,13). This way, the radiological protection ALARA principle will be efficiently applied.
Acknowledgments To the Núcleo de Imagens Médicas of Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS), for providing the film scanner and the calibration process developed for the correct digitization of images. To Professor Elaine Evani Streck, for her assistance, contributing in the research that originated the present study.
REFERENCES 1. Souza ABB, Dias SK, Mecca FA, et al. Avaliação do kerma no ar na superfície de entrada e da qualidade da imagem em radiografias de tórax. Radiol Bras. 2009;42:315-9. [ ] 2. Bontrager LK. Tratado de técnicas radiológicas e base anatômica. 7ª ed. Rio de Janeiro: Guanabara Koogan; 1999. [ ] 3. Rangayyan RM. Biomedical image analysis. Florida: CRC Press LLC; 2005. [ ] 4. ImageJ: image processing and analysis in Java. [acessado em 20 de novembro de 2008]. Disponível em: rsbweb.nih.gov/ij/ [ ] 5. Montgomery DC, Runger GC. Estatística aplicada e probabilidade para engenheiros. 2ª ed. Rio de Janeiro: LTC; 2003. [ ] 6. Silva RR. Processamento de imagens radiográficas e análise de defeitos em cordão-de-solda [dissertação de mestrado]. Rio de Janeiro: COPPE/UFRJ; 2000. [ ] 7. Universidade de Caxias do Sul. Centro de Ciências Exatas e Tecnologia. Departamento de Física e Química. Medidas e grandezas físicas E. Teoria de erros. [acessado em 5 de novembro de 2008]. Disponível em: http://hermes.ucs.br/ccet/defq/mlandreazza/Teo-ince.htm#_toc416523375 [ ] 8. Shinohara AH, Acioli E, Khoury HJ. Avaliação da técnica de radiografia digital em gamagrafia. Anais da 6ª COTEQ - Conferência sobre Tecnologia de Equipamentos. Salvador, BA, Brasil; agosto de 2002. [ ] 9. Oliveira DF. Análise da radiografia computadorizada em condições de águas profundas [dissertação de mestrado]. Rio de Janeiro: COPPE/UFRJ; 2007. [ ] 10. Sprawls P Jr. Physical principles of medical imaging. Madison: Medical Physics Pub; 1995. [ ] 11. Cranley K, Gilmore BJ, Fogarty GWA, et al. Catalogue of diagnostic X-ray spectra and other data. IPEM Report 78. York: Institute of Physics and Engineering in Medicine; 1997. [ ] 12. Brasil. Ministério da Saúde. Diretrizes de proteção radiológica em radiodiagnóstico médico e odontológico. Portaria nº 453, de 1º de junho de 1998. Brasília: Diário Oficial da União, 2 de junho de 1998. [ ] 13. Luz MR. Estudo comparativo de dose e qualidade da imagem em exames de tórax realizados em sistemas analógicos e digitalizados. Revista da Graduação. 2009;2:26-81. [acessado em 16 de junho de 2009]. Disponível em: http://revistaseletronicas.pucrs.br/ojs/index.php/graduacao/issue/view/229 [ ]
Mailing address: Received August 17, 2009.
* Study developed at Pontifícia Universidade Católica do Rio Grande do Sul (PUCRS) - Faculdade de Física, Porto Alegre, RS, Brazil. |