The dosimetric parameters of small diameter photon beams of a commercially available radiosurgery system were measured in a water phantom using a p-type Si photon diode. Tissue maximum ratios, total scattering factor and beam profiles at 5 and 10 cm depth were measured for 17 circular beams ranging from 5 mm to 50 mm in diameter, in 1.5 mm steps. The total scattering factor decreased slowly from 0.947 to 0.888 for the cones with diameter between 50 mm and 12.5 mm (7% variation) whereas for cones with diameter between 10 mm and 5 mm the factor decreased more steeply, from 0.854 to 0.666 (28% variation). These tissue maximum ratio data are consistent with the data published by other authors. The beam profiles measured in the x and y directions were aproximately 0.2 mm between the two directions for all cone sizes. The full widths at half maximum were within 1 mm of the nominal cone sizes.
Resumo:
Os parâmetros dosimétricos de um feixe de raios X de pequeno diâmetro para um sistema de radiocirurgia comercial foram medidos em água com um detector de diodo de Si do tipo p. As razões tecido-máximo, o fator de espalhamento total e os perfis dos feixes a profundidades de 5 e 10 cm foram medidos para 17 feixes de diâmetros circulares de 5 mm a 50 mm, em incrementos de 2,5 mm. Os fatores de espalhamento totais caíram lentamente, de 0,947 para 0,888 entre os cones de 50 mm e 12,5 mm de diâmetro (variação de 7%); para os cones entre 10 mm e 5 mm de diâmetro, esta queda foi bem maior, de 0,854 para 0,666 (variação de 28%). Os valores obtidos para a relação tecido-máximo são consistentes com dados publicados. Os perfis dos feixes foram medidos nas direções x e y, e estão dentro de 0,2 mm para todos os cones entre as duas direções. A medida da largura à meia-altura se encontra dentro de 1 mm com o diâmetro nominal dos cones.
|