Radiologia Brasileira - Publicação Científica Oficial do Colégio Brasileiro de Radiologia

AMB - Associação Médica Brasileira CNA - Comissão Nacional de Acreditação
Idioma/Language: Português Inglês

Vol. 34 nº 2 - Mar. / Apr.  of 2001

ORIGINAL ARTICLE
Print 

Page(s) 89 to 94

Quality control of noninvasive voltage meters in X-ray equipment, diagnostic level

Autho(rs): Renato Assenci Ros, Linda V.E. Caldas

PDF Português      

Texto em Português English Text

Keywords: X-radiation, Quality control, Noninvasive systems

Descritores: Radiação X, Controle de qualidade, Sistemas não-invasivos

Abstract:
We present the quality control tests performed on a monophasic X-ray equipment and on two noninvasive monitoring systems. The results were compared with the values obtained using an invasive monitoring system and a hyperpure germanium detector, providing correction factors for utilization of the noninvasive voltage meters in the quality control program of X-ray equipment. We concluded that there is a need for periodic calibration and control.

Resumo:
Este trabalho apresenta os testes de controle de qualidade feitos em um equipamento de raios X monofásico e em dois sistemas de monitoração não-invasivos. Os resultados foram comparados com os valores apresentados por um sistema de monitoração invasivo e por um detector de germânio hiperpuro, fornecendo fatores de correção para utilização dos medidores de tensão não-invasivos no programa de controle de qualidade de equipamentos de raios X, mostrando a necessidade de calibração e controle periódico.


 
RB RB RB
GN1© Copyright 2024 - All rights reserved to Colégio Brasileiro de Radiologia e Diagnóstico por Imagem
Av. Paulista, 37 - 7° andar - Conj. 71 - CEP 01311-902 - São Paulo - SP - Brazil - Phone: (11) 3372-4544 - Fax: (11) 3372-4554